Performance Evaluation of Hematology Analyser Sysmex KX – 21 for Platelet and WBC Count
DOI:
https://doi.org/10.21649/akemu.v18i3.420Abstract
Objective: To evaluate the performance of Sysmex KX – 21 for the estimation of Platelet and WBC count in comparison with microscopy.
Study Design: It is prospective cohort study.
Place and Duration: It was conducted in the Patho-logy department, Allama Iqbal Medical College, from 7th Oct. 2009 to 6th Nov. 2009.
Materials and Methods: Out of the total 1500 samp-les, 1458 (97%) samples after exclusions were finally evaluated for the study. On each sample, automated blood cell count was performed using Sysmex KX – 21 and the results were compared by peripheral smear examination for Platelet count, WBC count and its dif-ferential. The samples were from Medicine, Pediatrics and Oncology wards only.
Results: Out of 1458 samples, WBC count of 1452 (99.5%) samples matched the total count on peripheral smear examination. 06 (0.5%) samples showed spur-ious WBC count due to the presence of nucleatedDownloads
How to Cite
Issue
Section
License
This is an open-access journal and all the published articles / items are distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. For comments publications@kemu.edu.pk