Comparison of a Cellphone-Based Fundoscope with Traditional Handheld Direct Fundoscope for Checking Red Reflex in Neonates

This is an outdated version published on 10/27/2022. Read the most recent version.

Comparison of a Cellphone-Based Fundoscope with Traditional Handheld Direct Fundoscope for Checking Red Reflex in Neonates

Authors

  • Muhammad Tauseef Omer
  • Amna Rashid Butt
  • Tayyaba Khawar Butt
  • Muhammad Asif Siddiqui
  • Saba Bakht
  • Sulman Javaid

DOI:

https://doi.org/10.21649/akemu.v28iSpecial%20Issue%20(Jul,%20Sep).5198

Keywords:

Red reflex, direct fundoscopy, cellphone-based fundoscopy, cataract screening, red reflex screening

Abstract

Background: Traditional handheld fundoscopes are expensive instruments and require training and expertise for their use to check red reflex in neonates. Cellphone based fundoscopes may prove to be a cheaper alternative to check red reflex for screening of congenital cataracts in neonates. Objective: Evaluation of the diagnostic accuracy of a cellphone-based fundoscope Opti Card for screening of cataracts in neonates by checking pupillary red reflex. Methods: An instrument validation study was conducted in the Neonatology Department of Services Hospital Lahore for 3 months from 1st August to 30th October 2021 to assess the diagnostic accuracy of cellphone-based fundoscope OptiCard for screening of cataracts in neonates using Wilson Score. Results: Eyes of 110 neonates were examined for red reflex. Cellphone-based fundoscope OptiCard was found to have a sensitivity of 100% for diagnosing abnormal red reflex while its specificity was found to be 99.06% as compared to the traditional handheld direct fundoscope. Conclusion: Cellphone-based fundoscope OptiCard has good diagnostic accuracy for detecting abnormal red reflex and it allows the images of pupillary glow to be stored for a photo follow-up.

Downloads

Published

10/27/2022

Versions

How to Cite

Omer, M. T. ., Butt, A. R. ., Butt, T. K. ., Siddiqui, M. A. ., Bakht, S. ., & Javaid, S. . (2022). Comparison of a Cellphone-Based Fundoscope with Traditional Handheld Direct Fundoscope for Checking Red Reflex in Neonates. Annals of King Edward Medical University, 28(Special Issue (Jul, Sep), 354–359. https://doi.org/10.21649/akemu.v28iSpecial Issue (Jul, Sep).5198

Most read articles by the same author(s)

Similar Articles

<< < > >> 

You may also start an advanced similarity search for this article.

Loading...